Development of two-beam coal dust concentration and dispersion meter with temperature drifts compensation
Keywords
вимірювач
концентрація
вугільний пил
дисперсність
компенсація
температура
Abstract
Concentration and dispersion meter of coal dust is developed and studied with temperature drift compensation. In its la-boratory tests is set the basic absolute error of dust concentration measurement in the range from 0 to 3 g/m3, which does not exceed ±8 mg/m3, which is more than less a magnitude order of the dust concentration measurement absolute error of the aspiration measuring devices (±200 mg / m3). Additional dust concentration measurement error in the temperature range from 5 to +35 ○C is ±2,5 mg/m3 that no more than 35 % of the dust concentration measuring basic error and totally satisfies the meter operation conditions of the mine atmosphere.